Electron Microscope / Microscopy Rooms
Environmental Rooms Specialists - Airmont Inc.



July 29, 2010




Electron Microscope / Microscopy Rooms

Enviornmental Room Image

Enviornmental Room Image

Enviornmental Room Image

Air Conditioning - The air conditioning temperature must be adjustable between 19 degrees Celsius and 25 degrees Celsius. The air conditioning temperature regulation system shall be proportional control.

Temperature & Humidity Specification for Electron Microscope / Microscopy Room:
Recommended ambient temperature (for operator) 20 degrees Celsius
Maximum ambient temperature range for operator within specification 18 to 23 degrees Celsius
20-25% relative humidity at 20 degrees Celsius
Maximum permissible rate of temperature change for operation within specifications 0.5 degrees Celsius per hour temperature drift

Matters to Consider are:
A dust free, air-conditioned room is strongly recommended for the installation
An ambient noise level as low as possible
Good ventilation

All of Airmont's Electron Microscope/Microscopy Rooms can be typically retrofitted for an existing space







Home - About - Contact - Products - Environmental Room - Growth Room - Warm Room - Incubator Room - Insect Rearing Room - Tissue Culture Room - Cold Room - Laminar Air Flow Room - 4 Degree Celsius Room - Clean Room - Negative Pressure Isolation Room - Tuberculosis (TB) Isolation Room - Positive Pressure Isolation Room - Electron-Microscope / Microscopy Room - Controls -